Adopting a suspended frame design, it can cover a large area of illumination and can move and rotate in multiple dimensions,Meet the needs of various shooting positions. The electric control panel of the rack motion control panel is controllable and can be controlled freely; Efficient one click positioning, Realize the linkage control between the chest film holder and the hanging machine head, which is convenient for automatic rotation during upright and recumbent shooting,Convenient for users to use,Adopting an amorphous silicon flat panel detector, it has strong stability, better and clearer images.
specifications
I. High-frequency High-voltage Generator
1. Maximum Output Power: 50kW; 2. Input Power Supply: 380VAC three-phase; 3. Output Voltage: 40~150kV; 4. Tube Current Range: 10~800mA; 5. Exposure Time Range: 1ms-10000ms; 6. mAs Range: 0.1mAs~800mAs; 7. Support AEC and APR Automatic Exposure; 8. Fault self-diagnosis and display;;
II. X-ray Ball Tube Assembly
1. Focus Design: 0.6mm/1.2mm; 2. Nominal Electric Power: 22kW/54kW; 3. Maximum KV: 150kV; 4. Anode Target Angle: 12;
III.Filter Grid and Collimator
1. Grid material: aluminum-based grid; 2. Size: 18×18 inches (48×48cm) 3. Grid ratio of filter grid: 10:1; 4. Lines/inch: 215C; 5. Focal distance of filter grid: 1000MM; 6. Dedicated collimator with light field and multi-leafDR for line restriction
IV.Gantry
The lateral movement range of the overheadtrack suspension is not less than 2400 mm;
The lifting height of the detector center of thecolumn is 500 mm~1600 mm, with a deviation of no more than ±5%;
The rotational angle of the detector on the column is 0 ~90 degrees, with a deviation of no more than ±1%;
The horizontal travel of the X-ray tube assembly is not less than 2000 mm;
The lifting range of the X-ray tube assembly should be 1400 mm, with a deviation of no more than ±5%;
Rotation angle of X-ray tube assembly: 0 ~90 degrees, with a deviation of no more than ±1%;