-High frequency inverter type X-ray generator -Automatic exposure control -Microprocessor controlled operating system -Digital display of radiographic parameters e.g. KV, mA, time etc. -Memory for different radiographic examinations factors -Digital display of radiographic examinations factors -X-ray system is suitable for Bucky&General radiography -Anatomical programsare more than 200 kinds -Light beam collimator
specifications
I. High-frequency High-voltage Generator
1. Maximum Output Power: 50kW;: 2. Input Power Supply: 380VAC three-phase; 3. Output Voltage: 40~150kV; 4. Tube Current Range: 10~800mA; 5. Exposure Time Range: 1ms-10000ms; 6. mAs Range: ≥0.1mAs~1000mAs; 7. Support AEC and APR Automatic Exposure; 8. Fault self-diagnosis and display;
1. Vertical Movement: 1200mm; 2. Longitudinal Movement: 2000mm; 3. Transverse movement: ±135; 4. Rotation of X-ray tube around cross-arm: ±1350;
IV. Floating Top Bucky Table; 1. Four-way movable tabletop with oscillating grid Bucky; 2. Lateral movement range of bed surface: ±280 3. Longitudinal movement range of detector on bed panel: ±300 mm;
V. Vertical Chest X-ray StandX
Vertical movement range of the chest X-ray stand:1400±20 mm;
VI. AEC (Optional)
1. Number of detection fields: 3; 2. Tube voltage range: 40-150KV; 3. Light time range: 1ms-10s;
VII. Grid Filter and Collimator
1. Grid material: aluminum-based grid; 2. Size: 18×18 inches (48×48cm); 3. Grid ratio of the grid filter is 10:1; 4. Density: 215C; 5. Focal length of grid filter: 180cm;
VIII.Flat Panel Detector
1. Detector Type: TFT amorphous silicon flat paneldetector; 2. Detector Panel Size: 430mm x 430mm; 3. Effective Imaging Size of Detector: 17×17 inches; 4. Pixel Matrix for Detector Acquisition: ≥3072×3072; 5. Pixel Spacing of Detector: ≤139 um; 6. Spatial Resolution of Detector: 3.7LP/MM; 7. Pixel Gray Scale: 16Bit; 8. Image Acquisition and Transmission Time: wired 1s,wireless 3s;